multi X 2500 Fault removal
111
272 Auto-injector – syringe: incorrect size
Cause Remedy
Dosing volume and syringe size differ Adjust dosing volume and syringe size
273 Auto-injector – syringe: not filled properly
Cause Remedy
Syringe not drawn all the way to the stop Draw syringe completely
Insert syringe
Auto-injector: no connection
Auto-injector coupling was not found
Acknowledge message
Initialize
300 Temperature controller: no connection
Cause Remedy
Communication error Acknowledge message
Initialize
304 Temperature controller: communication error
Cause Remedy
Temperature cannot be set Acknowledge message
Initialize
Initialize
11.2 Equipment faults and analytical problems
11.2.1 General notes
Other problems not detected by the system monitoring can also occur. Starting a
measurement is possible. Such errors are usually detected on the basis of measuring
results which seem to be implausible (analytical problems) or they are clearly indicated
by the equipment technology.
If the suggested solutions are not successful, inform the customer service department
at Analytik Jena.
11.2.2 Equipment fault
Thermocouple connector not connected
Temperature set incorrectly in the software
No method loaded
Malfunction in power supply
Malfunction in the internal electronics
Connect thermocouple connector (4 in Fig. 5, p. 20)
Check temperature configuration in the method
Load method
Switch on the device,
check the internal fuse
check the connection b/w basic device and PC
Report to customer service