Fault removal multi X 2500
110
222 Sampler: wrong syringe size
Cause Remedy
No syringe inserted Insert syringe into autoX 112
Dosing volume in the method greater than inserted
syringe
Adjust dosing volume
The method for dosing liquids is to be enabled and the
gripper and solids track are still inserted
Insert sample rack for liquid samples into
autoX 112
Initialize
224 Sampler: wrong gripper
Cause Remedy
No gripper inserted Insert gripper into autoX 112
The method for solids is to be enabled and the syringe
and sample rack for liquid samples are still inserted
Insert solids rack into autoX 112
Initialize
225 Sampler: position not reached
Cause Remedy
autoX 36 – target position is not reached Check autosampler for mechanical hindrance,
remove hindrance if required
Replace rack
Initialize
226 Sampler: runtime exceeded
Cause Remedy
Message from autoX 112 sampler saying that the
movement is completed takes too long (autoX 112
defective)
Record log files
Report to customer service
Communication error after ABD was detected during
initialization
Initialize
232 Flame sensor error
Cause Remedy
Flame sensor calibration failed Record log files
Report to customer service
260 Sample handling missing
Cause Remedy
No sampler module detected Connect at least one sample introduction module
(autoX, auto-injector, ABD, MBD)
Initialize
Autoinjector - syringe: no connection
No communication with auto-injector
Initialize
271 Autoinjector - syringe: runtime exceeded
Cause Remedy
Communication error with auto-injector
Completion message at dosing end not received
Acknowledge message
Check auto-injector
Initialize