multi X 2500 Specification
123
Measurement mode Potentiometry
Measuring range 0.01 to 10 µg Cl
Generator current 100 µA
Electrolyte volume 65 ml
Measurement mode Bi-amperometry
Measuring range 1 to 100 µg Cl
Generator current 1 mA
Electrolyte volume 15 to 20 ml
Measurement mode Bi-amperometry
Measuring range 10 to 1000 µg Cl
Generator current 10 mA
Electrolyte volume 120 ml
Argon
For AOX horizontal, EOX, POX, TX
4.6 (free from halogens and hydrocarbons)
Primary pressure: 4 to 6 bar
Gas consumption: approx. 40 l/h
Oxygen 4.5 (or better)
Primary pressure: 4 to 6 bar
Gas consumption: approx. 10 l/h
Power supply 110 to 240 V AC, 50/60 Hz
Fuse for basic device T 10 A H (2x)
Fuse for chlorine module T 4.0 A H (2x)
Only use original fuses from Analytik Jena!
Typical average power consumption
of the basic device
1000 VA
Typical average power consumption
of the Cl module
100 VA
PC interface
USB
Temperature 20 to 35 °C
Humidity 90 % max at 30 °C
Air pressure 0.7 bar to 1.06 bar
Max. permissible altitude 2000 m
Temperature 15 to 55 °C
Humidity 10 to 30 % (use desiccant)
Measuring cell "high
sensitive"
Measuring cell "sensitive"
Measuring cell "high
concentration"
Gas supply
Electrical variables
Ambient conditions
during operation
Ambient conditions
for storage