multi X 2500 Installation and commissioning
47
1. Connect the mains cables to the mains connections of the
multi X 2500 and the chlorine module. Plug in the mains
cable to the mains power outlet.
2. Connect the multi X 2500 and the chlorine module using the
serial cable.
Cl-Coul interface on the rear of the chlorine module (1)
Cl-Coul interface on the rear of the basic device (2)
Standard version (AOX vertical):
3.
Connect the exhaust air hose to the outlet "exhaust" of the
chlorine module (3). Connect the hose to an extraction
system or place it into an exhaust vent.
4.
Connect the provided gas hose to the oxygen supply of the
laboratory and the gas connection O
2
IN on the rear of the
device.
Optional versions (AOX horizontal, EOX, POX, TX):
5.
Connect the exhaust air hose to the outlet "waste" of the
chlorine module (4). Connect the hose to an extraction
system or place it into an exhaust vent.
6.
Connect the provided gas hose to the oxygen supply of the
laboratory and the gas connection O
2
IN on the rear of the
device.
7.
Connect the provided gas hose to the argon supply of the
laboratory and the gas connection Ar IN on the rear of the
device.
The outlet "exhaust" on the basic device does not need to be connected to the
extraction system because this outlet is not used to expel exhaust air to the
surrounding air.