multi X 2500 EOX determination
81
Alternatively, keep the samples at hand for introducing them manually with a
syringe or an auto-injector.
11. Create a new method under M
ETHOD METHOD NEW and release it.
12. Activate the new method or an already existing method using the menu command
M
ETHOD METHOD ACTIVATE.
13. Starting the measurement:
Select S
TART START-ANALYSIS.
Select an analysis group or create a new one.
Create an analysis sequence.
Enter the sample ID for all samples in the field N
AME.
Confirm the entry with [OK].
Click the [S
TART MEASUREMENT] button.
The prepared analysis sequence is processed.
For manual sample introduction, follow the prompts in the control and analysis
software multiWin and introduce the samples one after the other.