multi X 2500 Maintenance and care
97
10. Hold the auto-protection assembly with the left hand and
pull the knob of the clamping mount with the right hand to
open the lock.
Remove the auto-protection assembly from the bracket on
the combustion furnace.
The condition of the filter in the auto-protection assembly must be checked every
three months. Proceed as follows:
1. Remove the auto-protection assembly.
2.
Check the filter for carbonization, soiling or cracks through
visual inspection.
If the filter is OK, re-install the auto-protection assembly.
If the filter needs replacing, follow the remaining instructions.
3. Unscrew the 4 screws attaching the pneumatic seal to the
auto-protection assembly.
4.
Remove the pneumatic seal from the auto-protection
assembly.
5. Remove the white intermediate ring.
Check/replace filter