Operation Cerabar M PMC51, PMP51, PMP55
24 Endress+Hauser
Menu path: Setup → Extended setup → Current output → Set URV
Set URV 014 Operation
The local display shows the parameter to be changed. The "mbar" unit is
defined in another parameter and cannot be changed here.
1
1 0 0 . 0 0 0
mbar
Press or to enter the edit mode.
The first digit is highlighted in black.
2
1 0 0 . 0 0 0
mbar
Use the key to change "1" to "5".
Press the key to confirm "5". The cursor jumps to the next position
(highlighted in black).
Confirm "0" with (second position).
3
5 0 0 . 0 0 0
mbar
The third digit is highlighted in black and can now be edited.
4
5 0 0 . 0 0 0
mbar
Use the key to change to the " " symbol.
Use to save the new value and exit edit mode. See next graphic.
5
5 0
¿
. 0 0 0
mbar
The new value for the upper range value is 50 mbar (0.75 psi).
Use to exit the edit mode for the parameter.
Use or to return to the edit mode.
6
5 0 . 0 0 0
mbar
6.4.4 Operating example: Accepting the pressure present
Example: Setting pos. zero adjustment.
Menu path: Main menu → Setup → Pos. zero adjust
Pos. zero adjust 007 Operation
1 Cancel The pressure for pos. zero adjustment is present at the device.
Confirm
2 Cancel Use or to switch to the "Confirm" option. The active option is highlighted in
black.
Confirm
3 Adjustment has
been accepted!
Use the key to accept the applied pressure as zero adjustment. The device
confirms the adjustment and goes back to the "Pos. zero adjust" parameter.
4 Cancel Use to exit the edit mode for the parameter.
Confirm