4.4 Installing the
Due to the orientation of the Cerabar M, there may be a shift in the zero point, i.e. when the
container is empty or partially full, the measured value does not display zero. You can
correct this zero point shift→  20"Funktion der Bedienelemente".
The local display can be rotated in 90° stages.
For PMP55, please refer to Section "Installation instructions for devices with diaphragm
seals – PMP55"→  13.
Endress+Hauser offers a mounting bracket for installing on pipes or walls.→  14,
Section "Wall and pipe mounting (optional)".
4.4.1 Installation instructions for devices without diaphragm seals – PMP51, PMC51
NOTICE
Damage to the device!
If a heated device is cooled during a cleaning process (e.g. by cold water), a vacuum develops
for a short time and, as a result, moisture can enter the sensor through the pressure
compensation element (1).
Mount the device with the pressure compensation element (1) pointing diagonally
downwards or to the side as much possible.
1
1
1
1
A0028471
Keep the pressure compensation and GORE-TEX® filter (1) free from contamination.
Cerabar M transmitters without diaphragm seals are mounted as per the norms for a
manometer (DIN EN 837-2). We recommend the use of shutoff devices and siphons. The
orientation depends on the measuring application.
Do not clean or touch process membranes with hard or pointed objects.
The device must be installed as follows in order to comply with the cleanability
requirements of the ASME-BPE (Part SD Cleanability):