Cerabar M PMC51, PMP51, PMP55 Commissioning
Endress+Hauser 31
Selection Confirm
Cancel
Factory setting Cancel
Pos. zero adjust (007) (gauge pressure sensors)
Write permission Operator/Maintenance/Expert
Description Pos. zero adjustment – the pressure difference between
zero (set point) and the measured pressure need not be
known.
Example Measured value = 2.2 mbar (0.033 psi)
You correct the measured value via the "Pos. zero adjust"
parameter with the "Confirm" option. This means that you
are assigning the value 0.0 to the pressure present.
Measured value (after pos. zero adjustment) = 0.0 mbar
The current value is also corrected.
Selection Confirm
Cancel
Factory setting Cancel
8.4 Configuring level measurement
See the Operating Instructions.
8.5 Configuring pressure measurement
See the Operating Instructions.
8.6 Backing up or duplicating the device data
See the Operating Instructions.