Deltabar S FOUNDATION Fieldbus Operation
Endress+Hauser 19
P01-xxxxxxxx-19-xx-xx-en-027
1. Press "+" or "–" to get to the editing mode.
2. The first digit is highlighted in black.
P01-xxxxxxxx-19-xx-xx-en-028
1. Use "+" to change "2" to "3".
2. Confirm "3" with "E". The cursor jumps to the next
position (highlighted in black).
P01-xxxxxxxx-19-xx-xx-en-029
The decimal point is highlighted in black, i.e. you can now
edit it.
P01-xxxxxxxx-19-xx-xx-en-030
1. Keep pressing "+" or "–" until "0" is displayed.
2. Confirm "0" with "E".
The cursor jumps to the next position. is displayed
and is highlighted in black. See next graphic.
P01-xxxxxxxx-19-xx-xx-en-031
Use "E" to save the new value and exit the editing mode.
See next graphic.
P01-xxxxxxxx-19-xx-xx-en-032
The new value for the damping is now 30.0 s.
– Jump to the next parameter with "E".
– You can get back to the editing mode with "+" or "–".
Onsite display Operation