Commissioning Deltabar S FOUNDATION Fieldbus
36 Endress+Hauser
The measured
variable is not in
direct proportion to
the measured
pressure as, for
example, with
containers with a
conical outlet. A
linearization table
must be entered for
the calibration.
LEVEL
SELECTION: Level
standard/
LEVEL MODE:
Pressure linearized
Via LINd.
MEASURAND
parameter:
Pressure + %
Pressure +
volume
Pressure + mass
Calibration with
reference
pressure:
semiautomatic
entry of
linearization
table, see
Operating
Instructions
BA00303P.
Calibration
without
reference
pressure:
manual entry of
linearization
table, see
Operating
Instructions
BA00303P.
Incorrect entries
are rejected by
the device
Customized
level, volume
and mass units
are possible
The measured
value display and
the TANK
CONTENT
parameter show
the measured
value.
–Two measured
variables are
required or
The container
shape is given
by value pairs,
such as height
and volume.
The 1st measured
variable %-height
or height must be
in direct proportion
to the measured
pressure. The 2nd
measured variable
volume, mass or %
must not be in
direct proportion to
the measured
pressure. A
linearization table
must be entered for
the 2nd measured
variable. The 2nd
measured variable
is assigned to the
1st measured
variable by means
of this table.
LEVEL
SELECTION: Level
standard/
LEVEL MODE:
Height linearized
Via COMB.
MEASURAND
parameter:
–Height +
volume
–Height + mass
–Height + %
–%-Height +
volume
–%-Height +
mass
–%-Height + %
Calibration with
reference
pressure: wet
calibration and
semiautomatic
entry of
linearization
table, see
Operating
Instructions
BA00303P.
Calibration
without
reference
pressure: dry
calibration and
manual entry of
linearization
table, see
Operating
Instructions
BA00303P.
Incorrect entries
are rejected by
the device
Customized
level, volume
and mass units
are possible
The measured
value display and
the TANK
CONTENT
parameter show
the 2nd measured
value (volume,
mass or %).
The LEVEL
BEFORE LIN
parameter displays
the 1st measured
value
(%-height or
height).
Measuring task LEVEL
SELECTION/LE
VEL MODE
Measured
variable options
Description Comment Measured value
display