Maintenance
64
7 Maintenance
The unit is maintenance-free.
7.1 Software update via the provided PC software
"
Caution!
When the program is updated, all measurement data in the memory is deleted. If the measured
values saved in the unit are still required, they should first be read out or the CompactFlash card
should be updated and removed from the unit. After the program transfer, all unit settings are reset
to the factory settings.
Procedure:
1. Start the supplied PC software
2. Connect the unit via USB to the PC (update only possible via USB connection!)
3. Select "Extras -> Special unit functions -> (select the unit) -> Transmit program"
4. Select interface parameters (Com-Port)
5. Select desired program file (*.prg) and confirm with OK.
7.2 Instruction for releasing a software option e.g.
"Integration + Analysis + Maths"
!
Note!
You need the provided PC software (at least version 1.23.0 or higher) for releasing a software
option.
The firmware of the unit must be at least version 2.00.00 or higher. Is this not the case, please
contact the manufacturer.
Please hold the release code, which you have received from the manufacturer, ready (see delivery
order).
It is necessary to read out the unit parameters before beginning. The unit's settings must be stored
in the PC software database.
Please make sure, that the unit is connected with a communication cable. The communication
port must be the same as used for the settings´ readout before (e.g. USB).
Procedure:
1. Start the supplied PC software
2. Select menu item "Extras -> Service -> Release options"
3. You will see a list of units stored in the PC database. Select the relevant unit.
4. Please enter the release code, which you have received from the manufacturer (see delivery
order), into the appearing pop-up window. Start the transmission of the release code with a
click on the button "OK".
5. If the transmission was successful a message on the provided PC software will appear. The unit
will reboot. A relevant message will be saved in the event list, indicating the successful option
release.
6. A new read out of the unit parameters is necessary.
The released option is now useable.