EasyManua.ls Logo

Endress+Hauser Ecograph T RSG30 - Ethernet and USB Interface Details

Endress+Hauser Ecograph T RSG30
86 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Wiring
17
4.2.6 "Ethernet" option
Ethernet connection
An IEEE 802.3 compatible connector on a shielded RJ45 plug on the rear side of the unit is available
as a network connection. Through this connection, the unit can be connected with a hub or switch
to units in an office environment. For safe spacing distances, the office equipment standard EN
60950 must be observed. The pin assignment corresponds to a standard MDI interface (AT&T258),
so that a shielded 1:1 cable with a maximum length of 100 meters (328 ft) can be used here. The
Ethernet interface is designed as 10BASE-T. Direct connection to a PC is possible with a crossover
cable. Half duplex and full-duplex data communication is supported.
Fig. 8: RJ45 socket (assignment AT&T256)
LED description
Beneath the Ethernet connection (see rear of unit) there are two light emitting diodes which
indicate the status of the Ethernet interface.
Yellow LED: link signal; is on when the unit is connected to a network. If this LED is not
illuminated then communication is impossible.
Green LED: Tx/Rx; when the unit is receiving or sending data it flashes irregularly, otherwise is
continuously on.
4.2.7 USB connection
USB connection
A compatible USB connection (V1.1, lowspeed 1.5 Mbit/s) is available on a shielded USB-B socket
at the front of the unit. The unit can be connected to a laptop via this connection. The pin
assignment corresponds to a standard USB interface, so that a shielded standard cable with a
maximum length of 3 meters (9.8 ft) can be used here.
!
Note!
USB2.0 is compatible to USB1.1, i.e. communication is possible.