EasyManua.ls Logo

Endress+Hauser iTEMP TMT142B - Mechanical construction

Endress+Hauser iTEMP TMT142B
100 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Technical data iTEMP TMT142B
58 Endress+Hauser
Interference immunity as per IEC/EN 61326 series, industrial requirements
Interference emission as per IEC/EN 61326 series, Class B equipment
A shielded cable that is grounded on both sides must be used for sensor cable lengths
of 30 m (98.4 ft) and more. The use of shielded sensor cables is generally
recommended.
Connection of the functional grounding may be needed for functional purposes.
Compliance with the electrical codes of individual countries is mandatory.
Overvoltage category II
Pollution degree 2
13.6 Mechanical construction
Design, dimensions Dimensions in mm (in)
132 (5.2)
135 (5.3)
112 (4.4)
106 (4.2)
121 (4.8)
121 (4.8)
!6.4
(0.25)
0
10
30
40
50
20
°F
A0025824
 17 Die-cast aluminum housing for general applications, or optional stainless steel housing (316L)
Electronics module and connection compartment
Display attachable in 90° stages
Weight Aluminum housing approx. 1.4 kg (3 lb), with display
Stainless steel housing approx. 4.2 kg (9.3 lb), with display
Materials
Housing Sensor terminals Nameplate
Die-cast aluminum housing AlSi10Mg/
AlSi12 with powder coating on
polyester base
Nickel-plated brass0.3 µm
gold flashed / cpl.,
corrosion-free
Aluminum AlMgl, anodized in black
316L 1.4404 (AISI 316L)
O-ring 88x3 HNBR 70° Shore PTFE
coating
- -

Table of Contents

Related product manuals