Input
Measured variable
The measured variable is the distance between the reference point and the product surface.
The level is calculated based on "E", the empty distance entered.
Optionally, the level can be converted to other variables (volume, mass) by linearization (32 points).
Measuring range
The following table describes the media groups and the possible measuring range as a function of
the media group.
Levelflex FMP53
Media group ε
r
Typical liquids
Measuring range
Rod probes
1 1.4 to 1.6 Liquefied gases, e.g. N
2
, CO
2
On request
2 1.6 to 1.9 Liquefied gas, e.g. propane
Solvents
Freon
Palm oil
One-piece: 4 m (13 ft)
Separable: 6 m (20 ft)
3 1.9 to 2.5 Mineral oils, fuels One-piece: 4 m (13 ft)
Separable: 6 m (20 ft)
4 2.5 to 4 Benzene, styrene, toluene
Furan
Naphthalene
One-piece: 4 m (13 ft)
Separable: 6 m (20 ft)
5 4 to 7 Chlorobenzene, chloroform
Nitrocellulose lacquers
Isocyanate, aniline
One-piece: 4 m (13 ft)
Separable: 6 m (20 ft)
6 > 7 Aqueous solutions
Alcohols
Acids, alkalis
One-piece: 4 m (13 ft)
Separable: 6 m (20 ft)
The formation of buildup, particularly of moist products, can reduce the maximum possible
measuring range.