Levelflex FMP51, FMP52, FMP54 PROFIBUS PA Commissioning (via operating menu)
Endress+Hauser 43
10.6 Configuration of an interface measurement
Only devices with the respective software option can be used for interface
measurements. This option is selected in the product structure: Feature 540 "Application
package", option EB "Interface measurement".
A0011177
13 Configuration parameters for interface measurements
R = Reference pioint of the measurement D
I
= Interface distance (Distance from reference point to lower medium)
E = Empty calibration (= zero point) L
I
= Interface
F = Full calibration (= span) D
L
= Distance
LN = Length of probe L
L
= Level
UP = Measured thickness upper layer
1. Setup → Tag description
Enter tag for measuring point.
2. Setup → Device address
Enter bus address of the device (only in case of software addressing).
3. Setup → Distance unit
Select distance unit.