Levelflex M FMP41C with HART/4...20 mA Operation
Endress + Hauser 33
5.3.2 Unlocking of configuration mode
If an attempt is made to change parameters when the instrument is locked, the user is automatically
requested to unlock the instrument:
"unlock parameter" (0A4):
By entering the unlock parameter (on the display or via communication)
100 = for HART devices
the Levelflex is released for operation.
Hardware unlock:
After pressing the
O
and
S
and
F
keys at the same time, the user is asked to enter the unlock
parameter
100 = for HART devices
"
Caution!
Changing certain parameters such as all sensor characteristics, for example, influences numerous
functions of the entire measuring system, particularly measuring accuracy. There is no need to
change these parameters under normal circumstances and consequently, they are protected by a
special code known only to the Endress+Hauser service organization. Please contact
Endress+Hauser if you have any questions.
O
and
S
and
F
press simultaneous
Please enter unlock code and confirm with