EasyManua.ls Logo

Endress+Hauser Levelflex M FMP41C - Page 34

Endress+Hauser Levelflex M FMP41C
92 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation Levelflex M FMP41C with HART/4...20 mA
34 Endress + Hauser
5.3.3 Factory settings (Reset)
"
Caution!
A reset sets the instrument back to the factory settings. This can lead to an impairment of the
measurement. Generally, you should perform a basic setup again following a reset.
A reset is only necessary:
if the instrument no longer functions
if the instrument must be moved from one measuring point to another
if the instrument is being de-installed /put into storage/installed
User input ("reset" (0A3)):
333 = customer parameters
333 = reset customer parameters
This reset is recommended whenever an instrument with an unknown 'history' is to be used in an
application:
The Levelflex is reset to the default values.
The customer specific tank map is not deleted.
The mapping can also be deleted in the "cust. tank map" (055) function of the "extended
calibr" (05) function group.
A linearisation is switched to "linear" although the table values are retained. The table can be
reactivated in the "linearisation" (04) function group.
List of functions that are affected by a reset:
A complete “basic setup" (00) must be activated.
ENDRESS + HAUSER
E
+
tank properties (002)
•medium cond. (003)
process proper. (004)
empty calibr. (005)
full calibr. (006)
output on alarm (010)
output on alarm (011)
outp. echo loss (012)
ramp %span/min (013)
delay time (014)
safety distance. (015)
in safety dist. (016)
overspill protection (018)
end of probe (030)
level/ullage (040)
linearisation (041)
customer unit (042)
max. scale (046)
diameter vessel (047)
check distance (051)
range of mapping (052)
start mapping (053)
offset (057)
output damping (058)
low output limit (062)
curr. output mode (063)
fixed cur. value (064)
•4mA value (068)
language (092)
back to home (093)
format display (094)
no of decimals (095)
sep. character (096)
unlock parameter (0A4)