Installation Liquiline CM442R/CM444R/CM448R
16 Endress+Hauser
Recommended installation direction: horizontal. The specified ambient conditions, in
particular the ambient temperatures, apply here only.
A vertical orientation is possible. However, in this case, additional fixing clips are required
at the place of installation to keep the device in position on the DIN rail.
Recommended installation of power unit for CM444R and CM448R: to the left of the device.
The following minimum distances must be observed:
Laterally from other devices incl. power units and from wall of cabinet:
at least 20 mm (0.79 inch)
above, below, in front of and behind the device (from cabinet door or other devices installed
there):
at least 50 mm (0.79 inch)
a0025369
Fig. 5: Minimum distances in mm (inch)