Liquiline CM442R/CM444R/CM448R Operation
Endress+Hauser 85
10.6 Additional functions
10.6.1 Limit switch
There are different ways of configuring a limit switch:
Assigning a switch-on and switch-off point
Assigning a switch-on and switch-off delay for a relay
Setting an alarm threshold and also outputting an error message
Starting a cleaning function
Operating mode
Function= "Diagnostics
message"
Options
as assigned
•Namur M
•Namur S
•Namur C
•Namur F
Factory setting
as assigned
as assigned
With this selection, the diagnosis messages are
transmitted over the binary output that you individually
allocated to it.
Namur M ... F
When you choose one of the Namur classes, all
messages allocated to that respective class are output.
You can also change the Namur class assignment for
every diagnostics message.
(Menu/Setup/General settings/Diagnostics/Device
behavior or Menu/Setup/Inputs/../Diagnostics
settings/Diag. behavior)
1) x:y = slot no. : input number
Path: Menu/Setup/Additional functions/Limit switches/Limit switches 1 to 8
Function Options Info
Source of data Options
•None
Connected sensors
Factory setting
None
Specify the input or output which should be the source of
data for the limit switch.
Measuring value Options
•None
Depends on theSource of
data
Factory setting
None
The measured value you can select depends on the
option selected under "Source of data".
Path: Menu/Setup/Outputs/Binary output x:y
1)
Function Options Info