Operation Liquiline CM442R/CM444R/CM448R
78 Endress+Hauser
10.5 Binary inputs and outputs
Hardware options, e.g. module "DIO" with 2 digital inputs and 2 digital outputs or fieldbus
module "485" enable the following:
via a digital input signal
measuring range switching for conductivity (upgrade code required, see accessories)
switching between different calibration datasets in the case of optical sensors
–an external hold
a cleaning interval to be triggered
a PID controller to be switched on and off, e.g. via the proximity switch of CCA250
the use of the input as an "analog input" for pulse-frequency modulation (PFM)
via a digital output signal
the static transmission (similar to a relay) of diagnostic states, point level switch states etc.
the dynamic transmission (comparable to a non-wearing "analog output") of PFM signals,
e.g. to control dosing pumps.