Electrical connection Liquiline CM442R/CM444R/CM448R
30 Endress+Hauser
6.4 Connecting additional inputs, outputs or relays
WARNING
!
Module not covered
No shock protection. Danger of electric shock!
If you are modifying or extending your hardware, always fill the slots from left to right. Do
not leave any gaps.
If not all the slots are occupied, always insert one or more dummy modules into the slot to
the right of the last module until all the positions are filled ( å 1, item 3). This ensures
the unit is shock-protected.
Always ensure shock protection is guaranteed particularly in the case of relay modules (2R,
4R, AOR).
The terminal strip ( å 23) is used to connect cable shields. Any shields that are
additionally needed must be connected to PE centrally in the cabinet via terminal blocks
provided by the customer.
a0016197
Fig. 26: Sensors with and without additional supply voltage
at sensor module 2DS