Liquiline CM442R/CM444R/CM448R Installation
Endress+Hauser 21
5.3 Post-installation check
1. Following installation, check all devices (controller, power unit, display) for damage.
2. Verify that the specified mounting distances have been observed.
3. Verify that all securing clips have been snapped into place and that the components are
securely positioned on the DIN rail.
4. Ensure that the temperature limits at the mounting location are observed.
a0019880
NOTICE
Installation errors
Damage e.g. to the cable, or malfunctions
Lay the cables in such a way that they cannot get
squashed e.g. when closing the cabinet door.
Verify that you have actually put the display cable into
the RJ45 socket in the base module and not, for
example, into the Ethernet socket of the (optional)
module 485. Otherwise, your display will not function.