Liquiline CM442R/CM444R/CM448R Operation
Endress+Hauser 57
10.2.3 Automatic hold
If a device-specific hold status is activated, any cleaning previously started is interrupted.
When a hold is active you can only start manual cleaning.
Time zone Options
•None
Choice of 35 time zones
Factory setting
None
If no time zone is selected, then Greenwich Mean Time
is used (London).
DST Options
•Off
•Europe
•USA
•Manual
Factory setting
Off
The controller adapts the summertime/normal time
changeover automatically if you choose European or
American daylight saving time.
Manual means that you can specify the start and end of
daylight saving time yourself. Here, two additional
submenus are displayed in which you specify the
changeover date and time.
Path: Menu/Setup/General settings/Automatic hold
Function Options Info
Device specific hold
Setup menu Options
Disabled
•Enabled
Factory setting
Disabled
Decide whether the device outputs should switch to the
defined hold state when the particular menu is opened.
Diagnostics menu
Calibration active Factory setting
Enabled
Hold release time 0 to 600 s
Factory setting
0 s
The hold status is maintained for the duration of the
delay time when you switch to the measuring mode.
Path: Menu/Setup/General settings/Date/Time
Function Options Info

Table of Contents