Operation Liquiline CM442R/CM444R/CM448R
88 Endress+Hauser
Limit value Settings
Depends on the measured
value
Operating mode="Above limit check" or "Below limit
check"
a0018080
Fig. 63: Exceeding (A) and undershooting (B) a limit value (without hysteresis and switch-on delay)
1 Limit value
2 Alarm range
t
1
,t
3
,t
5
No action
t
2
,t
4
An event is generated
If the measured values (MV) are increasing, the relay contact is closed when the on-value is exceeded ("Limit value" +
"Hysteresis") and the start delay has elapsed ("Start delay").
If the measured values are decreasing, the relay contact is reset when the off-value is undershot ("Limit value" -
"Hysteresis") and following the release delay ("Switch off delay").
Path: Menu/Setup/Additional functions/Limit switches/Limit switches 1 to 8