Ex specification PROFIBUS PA
and FOUNDATION Fieldbus
Wiring
Housing grounding Plastic housing
"
Caution!
You must connect the "Fixing plate" to the foundation ground with a separate functional ground line (2.5 mm
2
i14 AWG).
Stainless steel housing
"
Caution!
You must connect the outer ground connection of the housing to the foundation ground with a separate line
(GN/YE) (2.5 mm
2
i14 AWG).
Suitable for use in a FISCO system
Max. input voltage U
i
Max. input current I
i
Max. input P
i
Max. internal inductivity L
i
Max. internal capacity C
i
17.5 V
380 mA
5.32 W
negligible
negligible
a0003617
Housing grounding
1
2
Fixing plate
2.5 mm
2
(i14 AWG) functional ground
a0003616
Housing grounding
1
2
Outer ground connection
2.5 mm
2
(i14 AWG) line (GN/YE)