EasyManua.ls Logo

Endress+Hauser Liquiline M CM42 - Performance Characteristics; Ph; ORP; Conductivity

Endress+Hauser Liquiline M CM42
32 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
20
Performance characteristics
pH / ORP Reference temperature
25 °C (77 °F)
Measured value resolution
Maximum measured error
7)
Repeatability
Conductivity Reference temperature
25 °C (77 °F)
adjustable from –5 to 100 °C (23 to 212 °F)
8)
Measured value resolution
pH (glass electrodes and ISFET sensors) 0.01 pH
ORP: 1 mV
Temperature: 0.1 °C (0.1 °F)
SCS glass electrodes:
Glass:
Reference:
100 kΩ
100 Ω
Leak current, ISFET sensors: 100 nA
7) acc. to DIN IEC 746 part 1, under nominal operating conditions
pH (glass electrodes and ISFET sensors) 0.02 pH
ORP: 1 mV
Temperature:
NTC30k:
Pt100 / Pt1000:
0.5 K
0.25 K
SCS glass electrodes:
Glass:
Reference:
200 MΩ
200 Ω
Leak current, ISFET sensors: 100 nA
pH < 0.01
8) with software package "Advanced" or "Advanced Plus"
Conductivity < 0.1% from the measured value (4 digits)
minimum, at:
cell constant k 0.15 cm
-1
cell constant k 1.5 cm
-1
cell constant k > 1.5 cm
-1
0.001 μS/cm
0.01 μS/cm
0.1 μS/cm
Temperature (Pt100 / Pt1000) 0.01 °C (0.01 °F)
Resistivity
minimum
< 0.1% from the measured value (4 digits)
1 kΩ
.
cm

Other manuals for Endress+Hauser Liquiline M CM42

Related product manuals