13
Version 1: "Hard grounding"
Place the cable shield on the "Fixing plate".
Connect the cable wires as per the assignment.
Version 2: "Capacitive grounding"
Strip back the shield braiding, push the extension wire of the C module (pos. 1) onto the exposed shield and
fasten the clip:
Place the extension wire on the "Fixing plate".
Connect the cable wires as per the assignment.
Version 3: Fieldbus connection socket (accessories):
Screw the fieldbus connection socket (accessories) into the housing bushing.
Trim the connection cores of the socket to approx. 15 cm (5.9 ").
Connect the cable cores as per the assignment. In doing so, you must place the cable shield (GN/YE) on the
"Fixing plate".
a0004060
View in device (CPU module)
a0001640
Wiring diagram
a0004071
View in device (CPU module)
a0004073
Wiring diagram