CM42-MAA101EBE00 Diagnostics, troubleshooting
Endress+Hauser 19
101 Ref. impedance alarm F
Check reference electrode for contamination and damage
Clean reference electrode
102 Glass imp. too low alarm F
Impedance of glass membrane too low
Check pH sensor, replace if necessary
103 Ref. imp. too low alarm F
Reference impedance too low
Check reference electrode for contamination/clogging
Replace reference or combined electrode
104 Sensor supply bad F
Sensor operating voltage fluctuating
Check connection
Replace sensor cable or sensor
106 Glass impedance warning M
Check glass electrode for breakage and hairline cracks
Check medium temperature
Check electrode plug-in head for moisture and dry if necessary
107 Ref. impedance warning M
Check reference electrode for contamination and damage
Clean reference electrode
111 Glass imp. too low warning M
Check glass electrode for breakage and hairline cracks
Check medium temperature
Check electrode plug-in head for moisture and dry if necessary
112 Ref. imp. too low warning M
Check reference electrode for contamination and damage
Clean reference electrode
119 Temp offset lower limit F Check the temperature sensor of the sensor
Clean the sensor and recalibrate
Replace sensor
120 Temp offset lower limit F
127 Invalid TAG group F
Use a sensor with a suitable sensor designation or sensor group.
128 Invalid TAG F
129 Sensor change aborted C Restart the sensor change
130 Calibration active C Wait for the calibration to finish.
131 PV not stable M
Sensor too old
Cable or connector defective
132 Temperature not stable M
134 Zero pnt. too high alarm M
Sensor old or defective
Diaphragm blocked
Buffer solutions too old or contaminated
135 Zero pnt. too high warning M
136 Zero pnt. too low warning M
137 Zero pnt. too low alarm M
138 Slope too low alarm M
139 Slope too low warning M
No. Display text Cat. Tests and/or remedial measures

Other manuals for Endress+Hauser Liquiline M CM42