Diagnostics, troubleshooting CM42-MAA101EBE00
20 Endress+Hauser
172 Operating time alarm M
The operating time for the sensor which you specified has been reached.
Replace the sensor to ensure the operability of the measuring point. If a
new sensor is used, the counters are reset to the new sensor data.
173 Oper.time >80°C alarm M
The sensor has reached the operating time under extreme conditions
which you specified. Replace the sensor to ensure the operability of the
measuring point. If a new sensor is used, the counters are reset to the
new sensor data.
174 Oper.time >100°C alarm M
175 Oper.time <-300mV alarm M
176 Oper.time >300mV alarm M
177 Delta slope alarm M
Sensor aged. Replace the sensor.
178 Delta zero alarm M
180 Cal. expired alarm M
The operated hours counter has reached its limit. Perform a calibration.
This resets the counter.
181 No. steril. alarm M
The number of sterilizations which you specified has been reached.
Replace the sensor.
182 Operating time warning M
The operating time for the sensor which you specified has almost been
reached. Prepare to replace the sensor.
183 Oper.time >80°C warn M
The sensor will soon have reached the operating time under extreme
conditions which you specified. Prepare to replace the sensor.
184 Oper.time >100°C warn M
185 Oper.time <-300mV warn M
186 Oper.time >300mV warn M
190 Cal. expired warning M The operated hours counter is reaching its limit. Plan a calibration.
191 No. steril. warning M The number of sterilizations which you specified will be reached shortly.
203 Wrong transmitter type F
Use a sensor module that suits the software.
(Software version 10... for pH, 13... for conductivity, 20... for oxygen)
215 Simulation active C Active corresponding to your settings
218 Current output not available F Contact the Service Team!
219 Power supply bad C
Interference amplitude of power supply too high
Incorrect power supply or unshielded cables used
Connect the device to a clean power supply.
Make sure the cable is grounded correctly.
The error is not rectified until the interference amplitude is gone or if
it is in the frequency range for HART communication.
238 Delta slope alarm C
Sensor aged. Replace the sensor.
239 Delta zero alarm C
318 Glass imp. too high alarm M
Sensor check system warning
Check glass electrode for breakage and hairline cracks
Check medium temperature
319 Ref. imp. too high alarm M
Sensor check system warning
Check reference electrode for contamination and damage
Clean reference electrode
No. Display text Cat. Tests and/or remedial measures