Liquiphant FTL64
26 Endress+Hauser
Humidity
Operation up to 100 %. Do not open in a condensing atmosphere.
Operating altitude
As per IEC 61010-1 Ed.3:
Up to 2 000 m (6 600 ft) above sea level
Can be extended to 3 000 m (9 800 ft) above sea level if overvoltage protection is used
Climate class
As per IEC 60068-2-38 test Z/AD
Degree of protection
In accordance with DIN EN 60529, NEMA 250
IP66/IP68 NEMA 4x/6P
Types of housing:
Single compartment; plastic
Single compartment; aluminum, coated; Ex d/XP
Single compartment; 316L, cast; Ex d/XP
Dual compartment L-shaped, aluminum, coated; Ex d/XP
Ordering information: Select the required option in the "Electrical connection" order code.
Exclusion criteria are taken into account automatically.
If the "M12 plug" option is selected as electrical connection, then IP66/67 NEMA TYPE 4x is
valid for all housing types.
Vibration resistance
As per IEC60068-2-64-2009
a(RMS) = 50 m/s
2
, f = 5 to 2 000 Hz, t = 3 planes x 2 h
Shock resistance
In accordance with IEC60068-2-27-2008: 300 m/s² [= 30 g
n
] + 18 ms
g
n
: standard acceleration of gravity
Mechanical load
Support the device in the event of severe dynamic load. Maximum lateral loading capacity of the pipe
extensions and sensors: 75 Nm (55 lbf ft).
 Additional details in the "Support the device" section.
Electromagnetic
compatibility
Electromagnetic compatibility as per EN 61326 series and NAMUR recommendation EMC (NE21)
The requirements of EN 61326-3-1 for the safety function (SIL) are fulfilled
 Details are available in the supplementary Functional Safety Manual.
Process
Process temperature range
–60 to +230 °C (–76 to +446 °F)
–60 to +280 °C (–76 to +536 °F)
–50 to +230 °C (–58 to +446 °F) with PFA coating (conductive)
300 °C (572 °F) for a maximum of 50 h (cumulative)
Pay attention to the pressure and temperature dependence, see the "Sensor process pressure range"
section.
Thermal shock
Without restrictions within the process temperature range.
With PFA coating (conductive): ≤ 120 K/s
Process pressure range
L
WARNING
The maximum pressure for the measuring device is dependent on the lowest-rated element,
with regard to pressure, of the selected components. This means that it is necessary to pay
attention to the process connection as well as the sensor.
For pressure specifications, see the "Mechanical construction" section.
The measuring device must be operated only within the specified limits!
The Pressure Equipment Directive (2014/68/EU) uses the abbreviation "PS". The abbreviation "PS"
corresponds to the MWP (maximum working pressure) of the measuring device.