Liquistation CSF33 Diagnostics and troubleshooting
Endress+Hauser 131
No. Message Factory settings Tests or remedial action
S
1)
D
2)
F
3)
241 Firmware failure F On On Internal device error
1. Update the software
2. Contact the Service Department
3. Replace the backplane (Service)
242 Firmware incomp. F On On
243 Firmware failure F On On
261 Electronics module F On On Electronics module defective
1. Replace the module
2. Contact the Service Department
262 Module connection F On On Electronics module not communicating
1. Check the cable connection , replace it if
necessary
2. Check the power supply of the sampling
control module
3. Contact the Service Department
263 Incomp. detected F On On Wrong kind of electronics module
1. Replace the module
2. Contact the Service Department
284 Firmware update M On Off Update completed successfully
285 Update error F On On Firmware update failed
1. Repeat
2. SD card error → use another card
3. Incorrect firmware → repeat with suitable
firmware
4. Contact the Service Department
302 Battery low M On Off Buffer battery of real time clock is low
The date and time are lost if the power is
interrupted.
Contact the Service Department (battery
replacement)
304 Module data F On On At least 1 module has incorrect configuration data
1. Check the system information
2. Contact the Service Department
305 Power consumption F On On Total power consumption too high
1. Check installation
2. Remove sensors/modules
306 Software error F On On Internal firmware error
Contact the Service Department
310 Temperature sensor F On On Temperature sensor PT1 in the climate control
module for sample compartment measurement is
defective
No temperature regulation possible for the
sample compartment
Unable to cancel the sampling program
Contact the Service Department
311 Temperature sensor F On On Temperature sensor PT2 in the sample
compartment is defective
No sample temperature measurement possible
An in-progress sampling program cannot be
canceled
Replace sensor