Technical data Liquistation CSF33
174 Endress+Hauser
Permitted cable diameter:
M20x1.5 mm: 7 to 13 mm (0.28 to 0.51")
M25x1.5 mm: 9 to 17 mm (0.20 to 0.67")
Mains fuse Optional fuses:
T3.15A (for 230V power supply)
Power consumption Version with vacuum pump: 290 VA
Version with peristaltic pump: 290 VA
Power failure
Real-time clock: lithium battery, type CR2032
15.6 Performance characteristics
Sampling methods Vacuum pump/peristaltic pump/sampling assembly:
Event sampling
Single and multiple samples
Sampling table
Vacuum pump:
Time-paced
In proportion to volume
Peristaltic pump:
Time-paced
In proportion to volume
Flow proportional sampling/time override (CTVV)
Dosing volume Vacuum pump:
20 to 350 ml (0.7 to 12 fl.oz.)
Peristaltic pump:
10 to 10000 ml (0.3 to 340 fl.oz.)
The dosing accuracy and the repeatability of a sample volume < 20 ml can vary,
depending on the specific application.
Dosing accuracy Vacuum pump:
± 5 ml (0.17 fl.oz.) or 5 % of the set volume
Peristaltic pump:
± 5 ml (0.17 fl.oz.) or 5 % of the set volume
Repeatability 5 %
Intake speed > 0.5 m/s (> 1.6 ft/s) for ≤ 13 mm (1/2") ID, as per EN 25667, ISO 5667, CEN 16479-1
> 0.6 m/s (> 1.9 ft/s) for 10 mm (3/8") ID, in accordance with Ö 5893; US EPA
Suction height Vacuum pump:
Max. 6 m (20 ft) or max. 8 m (26 ft), depending on the version
Peristaltic pump:
Max. 8 m (26 ft)