EasyManua.ls Logo

Endress+Hauser Liquistation CSF33 - Overview of Diagnostic Information

Endress+Hauser Liquistation CSF33
180 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Diagnostics and troubleshooting Liquistation CSF33
130 Endress+Hauser
Menu/Setup/Run the ../Extended setup/Diagnostics settings/Diag. behavior
Function Options Info
Diagnostic message Selection
Off
On
Factory setting
Depends on the
Diag. code
You can deactivate or reactivate a diagnostic
message here.
Deactivating means:
No error message in the measuring mode
No error current at the current output
Error current Selection
Off
On
Factory setting
Depends on the
Diag. code
Decide whether an error current should be
output at the current output if the diagnostic
message display is activated.
In the event of general device errors, the error
current is switched to all the current outputs. In
the event of channel-specific errors, the error
current is only switched to the current output in
question.
Status signal Selection
Maintenance (M)
Out of specification (S)
Function check (C)
Failure (F)
Factory setting
Depends on the
Diag. code
The messages are divided into different error
categories in accordance with NAMUR NE 107.
Decide whether you want to change a status
signal assignment for your application.
Diag. output Selection
None
Binary outputs
Alarm relay
Relay
Factory setting
None
You can use this function to select a relay output
and/or binary output to which the diagnostic
message should be assigned.
An alarm relay is always available,
regardless of the device version. Other
relays are optional.
Before you can assign the message to an output:
Configure one of the output types mentioned as
follows:
Menu/Setup/Outputs/(Alarm relay or Binary
output or relay)/Function = Diagnostics and
Operating mode = as assigned.
Cleaning program Selection
None
Cleaning 1 ... 4
Factory setting
None
Decide whether the diagnostic message should
trigger a cleaning program.
You can define cleaning programs under:
Menu/Setup/Additional functions/Cleaning.
Detail information Read only Here you can find more information on the
diagnostic message and instructions on how to
resolve the problem.
11.4 Overview of diagnostic information
11.4.1 Device-specific, general diagnostic messages
No. Message Factory settings Tests or remedial action
S
1)
D
2)
F
3)
202 Selftest active F On Off Wait for self-test to be finished
216 Hold active C On Off Output values and status of the channel are on
hold

Table of Contents

Other manuals for Endress+Hauser Liquistation CSF33

Related product manuals