Liquistation CSF33 Diagnostics and troubleshooting
Endress+Hauser 149
11.10 Device test
Menu/Diagnostics/Systemtest
Function Options Info
Power Supply Read only The current supply voltage is displayed.
Cooling system (only for version with sample compartment temperature regulation)
Check cooling
Power supply Read only The current supply voltage is displayed.
With AC power supply: 24 V ±0.5 V
With DC power supply: 22 to 28 V
Overcurrent Read only No: no error
Yes: the fan in the climate control module is
defective
> Contact the Service Department
Sample compartment Read only The current temperature of the sample
compartment is displayed.
Sample compartment Read only When you start the cooling test, the temperature
at the start time is displayed
Cooling test off or Cooling test on -> progress is displayed
Start Test Action Start the cooling test.
Stop test Action End the cooling test.
Check heating
Power supply Read only The current supply voltage is displayed.
With AC power supply: 24 V ±0.5 V
With DC power supply: 22 to 28 V
Overcurrent Read only No: no error
Yes: the heating is defective
> Contact the Service Department
Overcurrent Read only No: no error
Yes: the fan is defective
> Contact the Service Department
Sample compartment Read only The current temperature of the sample
compartment is displayed.
Sample compartment Read only When you start the heating test, the
temperature at the start time is displayed
Heating test off or Heating test on -> progress is displayed
Start Test Action Start the heating test.
Stop test Action End the heating test.
Manual sampling
Bottle configuration Read only
Bottle configuration Read only
Bottle configuration Selection
Front
Bottle 1
...
Back
Select which bottle should be filled with the
sample.
Sample volume 50 to 2000 ml
10 to 10000 ml
Factory setting
100 ml
You can change the sample volume in the
version with the peristaltic pump. The sample
volume can be changed.
Sample volume Factory setting
200 ml
The sample volume is preset at the factory in the
version with the vacuum pump.