Diagnostics and troubleshooting Liquistation CSF33
148 Endress+Hauser
DIAG/Simulation
Function Options Info
Current output x:y Simulation of an output current
This menu appears once for each current output.
Simulation Selection
Off
On
Factory setting
Off
If you simulate the value at the current output,
this is indicated on the display by a simulation
icon in front of the current value.
Current 2.4 to 23.0 mA
Factory setting
4 mA
Set the desired simulation value.
Alarm relay
Relay x:y
Simulation of a relay state
This menu appears once for each relay.
Simulation Selection
Off
On
Factory setting
Off
If you simulate the relay state, this is indicated
on the display by a simulation icon in front of
the relay display.
State Selection
Low
High
Factory setting
Low
Set the desired simulation value.
The relay switches in accordance with your
setting when you switch on the simulation. In
the measured value display you see On (= Low)
or Off(= High) for the simulated relay state.
Meas. inputs Simulation of a measured value (only for
sensors)
This menu appears once for each measuring
input.
Channel : parameter
Simulation Selection
Off
On
Factory setting
Off
If you simulate the measured value, this is
indicated on the display by a simulation icon in
front of the measured value.
Main value Depends on the sensor
Set the desired simulation value.
Sim. temperature Selection
Off
On
Factory setting
Off
If you simulate the temperature measured value,
this is indicated on the display by a simulation
icon in front of the temperature.
Temperature -50.0 to +250.0 °C
(-58.0 to 482.0 °F)
Factory setting
20.0 °C (68.0 °F)
Set the desired simulation value.