Diagnostics and troubleshooting Liquistation CSF33
148 Endress+Hauser
DIAG/Simulation
Function Options Info
Current output x:y Simulation of an output current
This menu appears once for each current output.
Simulation Selection
• Off
• On
Factory setting
Off
If you simulate the value at the current output,
this is indicated on the display by a simulation
icon in front of the current value.
Current 2.4 to 23.0 mA
Factory setting
4 mA
‣
Set the desired simulation value.
Alarm relay
Relay x:y
Simulation of a relay state
This menu appears once for each relay.
Simulation Selection
• Off
• On
Factory setting
Off
If you simulate the relay state, this is indicated
on the display by a simulation icon in front of
the relay display.
State Selection
• Low
• High
Factory setting
Low
‣
Set the desired simulation value.
The relay switches in accordance with your
setting when you switch on the simulation. In
the measured value display you see On (= Low)
or Off(= High) for the simulated relay state.
Meas. inputs Simulation of a measured value (only for
sensors)
This menu appears once for each measuring
input.
Channel : parameter
Simulation Selection
• Off
• On
Factory setting
Off
If you simulate the measured value, this is
indicated on the display by a simulation icon in
front of the measured value.
Main value Depends on the sensor
‣
Set the desired simulation value.
Sim. temperature Selection
• Off
• On
Factory setting
Off
If you simulate the temperature measured value,
this is indicated on the display by a simulation
icon in front of the temperature.
Temperature -50.0 to +250.0 °C
(-58.0 to 482.0 °F)
Factory setting
20.0 °C (68.0 °F)
‣
Set the desired simulation value.