Diagnostics and troubleshooting Liquistation CSF33
150 Endress+Hauser
Menu/Diagnostics/Systemtest
Function Options Info
Start sampling Action
Peristaltic pump
Pump purge Action
Pump purge, to stop
press ESC
Read only
Current pump
operating time
Read only
Power supply Read only The current supply voltage is displayed.
With AC power supply: 24 V ±0.5 V
With DC power supply: 22 to 28 V
Motor current Read only The current consumption of the pump is
displayed.
Vacuum Read only The vacuum is an indicator of the suction height.
-> 100 mbar corresponds to approx. 1 m suction
height
Medium detected Read only Yes: the medium was detected
No: no medium was detected
Pump suction Action
Pump suction, to stop
press ESC
Read only
Current pump
operating time
Read only
Power supply Read only The current supply voltage is displayed.
With AC power supply: 24 V ±0.5 V
With DC power supply: 22 to 28 V
Motor current Read only The current consumption of the pump is
displayed.
Vacuum Read only The vacuum is an indicator of the suction height.
-> 100 mbar corresponds to approx. 1 m suction
height
Medium detected Read only Yes: the medium was detected
No: no medium was detected
Vacuum pump
(only for version with
vacuum pump)
Action
Bottle configuration Read only
Bottle volume Read only
Distributor position Selection
Front
Bottle 1
...
Back
Select which bottle should be filled with the
sample.
Sample volume Factory setting
200 ml
The sample volume is preset at the factory.
Start sampling Action Perform sampling manually.
Progress Read only The progress of the sampling operation is
displayed.
Power supply Read only The current supply voltage is displayed.
With AC power supply: 24 V ±0.5 V
With DC power supply: 22 to 28 V
Motor current Read only The current consumption of the pump is
displayed.