Maintenance Liquistation CSF33
158 Endress+Hauser
1
2
6
3
4
5
A0013896
 38 Vacuum pump
1 Outflow hose
2 Dosing glass
3 Dosing chamber cover
4 Air hose connection
5 Lock for intake hose
6 Thread adapter nut for intake hose
1. Check the sample volume set under Menu/Setup/General settings/Sampling/Dosing
volume.
2. Release the thread adapter nut on the intake hose (item 6).
3. Turn the intake hose at the lock (item 5) to the "open" position and pull the hose
upwards to disconnect it.
4. Release the air hose (item 4) and remove the dosing chamber (item 2) from the front
along with the outflow hose (item 1).
5. Open the bayonet lock (item 3) and open the dosing chamber.
Dosing
1
2
3
A0014128
 39 Vacuum pump
1 Dosing tube
2 Allen screw
3 Air hose connection
1. Release the 2mm Allen screw with the key provided.
2. Set the sample volume by adjusting the dosing tube. Secure the dosing tube with the
screw.
3. Use the white scale (A) to dose without pressure and the blue scale (B) to dose with
pressure.