Liquistation CSF33 Maintenance
Endress+Hauser 165
3. Remove the pump tube.
4. Remove any silicone deposits on the roller and the flexible pump bracket.
5. Make sure the roller turns smoothly and evenly
Cleaning the distribution arm
Clean the distribution arm as follows:
1
2
3
A0014112
 46 Sample compartment
1
2
3
Distribution arm
motor
Drain pipe
Distribution arm
1. Release the outlet pipe (item 2).
2. Push up the splash guard.
3. Remove the distribution arm from the front.
4. Remove the cover.
5. Clean these parts with water or soapsuds. Use a bottle brush if necessary.
6. Reinstall the cleaned parts in reverse order.
Make sure the distribution arm is seated correctly! The distribution arm must be
locked as otherwise the rotation movement could be blocked or the system might no
longer approach the bottles correctly.
12.4.3 Sample compartment
The sample compartment has a continuous inner plastic lining.
1. Remove the bottle trays and the distribution pan.
2. Remove the bottles.
3. Remove the distribution arm. See also
4. Spray-clean the sample compartment with a water hose.
You can wash the PE and glass bottles in a dishwasher at 60 °C.