Electrical connection Liquistation CSF33
24 Endress+Hauser
A0013237
13 Terminal assignment
1 Assignment: 100 to 120 V/200 to 240 V AC ±10 %
2 Assignment: 24 V DC +15/-9 %
3 Rechargeable batteries (optional)
4 Internal 24 V voltage
5 Fuses (only for batteries)
6.1.7 Cable terminals
After connection, make sure that every cable end is securely in place. Terminated
cable ends, in particular, tend to come loose easily if they have not been correctly
inserted as far as the limit stop.
14 Press the screwdriver
against the clip (opens the
terminal)
15 Insert the cable until the
limit stop
16 Remove the screwdriver
(closes the terminal)