Electrical connection Liquistation CSF33
26 Endress+Hauser
6.2.3 Sampler controller
The connections for the sampler controller are located in the controller housing (→  25).
Analog inputs and binary inputs/outputs
125
123
124
225
223
224
191
192
291
292
145
146
245
246
1
2
A0012988
 18 Position of the terminals
1 Analog inputs 1 and 2
2 Binary inputs/outputs
Analog inputs
223
225
124
123125
224
Out
In
Gnd
Gnd
In
Out
*
**
*
**
1
2
A0012989
 19 Assignment of analog inputs 1 and 2
* Analog input for passive devices (two-wire transmitter), Out + In terminals (125/123 or 225/223)
** Analog input for active devices (four-wire transmitter), In + Gnd terminals (123/124 or 223/224)
CSXxx
125
123
+
-
133
134
CMxx
A0028652
 20 With two-wire transmitter, e.g. Liquiline M CM42