Commissioning Liquistation CSF33
40 Endress+Hauser
Manual sampling
15000 ml
MODE
1
Menu/Manual sampling
Distribution position
Bottle 1
Multiplier
A0036865-EN
1. Manual sampling is triggered by the MAN soft key. This pauses any program
currently running.
The current bottle configuration and the current sample volume are displayed.
You can select the distributor position. In peristaltic systems, you can also change
the sample volume.
In vacuum systems, Multiplier a multiple of a single manual sample can be taken
under . Specification of Multiplierrange of adjustment 1 to 50.
2. Select Start sampling
A new screen is displayed indicating the progress of the sampling process.
3. After manual sampling, a running program can be displayed and continued with the
ESC ESC button.
The sample volume for "Manual sampling" is not taken into account in the
calculated bottle volumes.
Programming for automatic sampling
Create a simple sampling program in the general overview under Select sampling
program/New/Basic or in the menu Menu/Setup/Sampling programs/Setup
program/New/Basic :
1. Enter the "Program name".
2. The settings from the Basic setup for bottle configuration and bottle volume are
displayed.
3. Sampling mode=Time paced CTCV is preset.
4. Enter the Sampling interval .
5. Enter the Sampling volume per sample. (For version with vacuum pump, configure
under Menu/Setup/General settings/Sampling .)
6. Select the Bottle change mode after number of samples or time for average samples.
With the option "Bottle change after a time", you can enter the change time and bottle
synchronization (None, 1st bottle change time, 1st time of change + bottle number).
The description for this can be found in the "Bottle synchronization" section.
With the option "Bottle change after a time", you can choose the bottle
synchronization before the start condition (None, 1st bottle change time, 1st time of
change + bottle number). The description for this can be found in the "Bottle
synchronization" section.
1. For Multiple bottles enter the number of bottles the sample should be distributed
over.
2. Start condition: immediately or after date/time