Liquisys M CLM223/253 Commissioning
Endress+Hauser 33
C C1 C111
C121
C131
C132 C133
A0027502
 21 Function code
To make it easier for you to select and find function
groups and functions, a code for the corresponding
field is displayed for each function →  20
The structure of this code is illustrated in →  21.
The function groups are indicated as letters in the first
column (see the names of the function groups). The
functions of the individual groups are displayed
incrementally by row and by column.
Factory settings
The first time the device is switched on, the factory setting is set for all the functions. The
table below provides an overview of the most important settings.
All other factory settings can be found in the description of the individual function groups
in the "System configuration" section (the factory setting is highlighted in bold).
Function Factory setting
Type of measurement Conductive measurement of conductivity,
Temperature measurement in °C
Type of temperature compensation Linear with reference temperature 25 °C (77 °F)
Temperature compensation Automatic (ATC on)
Limit value for controller 1 9999 mS/cm
Limit value for controller 2 9999 mS/cm
Hold Active during configuration and calibration
Measuring range 0 μS/cm to 2000 mS/cm (no measuring ranges to be
configured).
The setting is flowing and depends on the sensors
connected.
Current outputs 1 and 2* 4 to 20 mA
Current output 1: measured value for 4 mA signal
current
0 μS/cm
Current output 1: measured value for 20 mA signal
current
2000 mS/cm
Current output 2: temperature value for 4 mA signal
current*
-35.0 °C (-31 °F)
Current output 2: temperature value for 20 mA
signal current*
250.0 °C (482 °F)
* with appropriate version