EasyManua.ls Logo

Endress+Hauser Liquisys M CLM223 - Relay Configuration

Endress+Hauser Liquisys M CLM223
122 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Liquisys M CLM223/253 Commissioning
Endress+Hauser 53
Coding Field Range of
adjustment
(factory settings
in bold)
Display Info
P7 Enter maximum
permissible
duration for lower
CC setpoint limit
violation (field
P9)
60 min
0 to 2000 min
P7
min
Tmax Low
60
A0009053-EN
Only if P6=CC or AC CC.
P8 Enter maximum
permissible
duration for upper
CC setpoint limit
violation (field
P9)
120 min
0 to 2000 min
P8
min
Tmax High
120
A0009054-EN
Only if P6=CC or AC CC.
P9 Enter CC set point
(for P7/P8)
1000 µS/cm
0 to 9999 mS/cm
P9
µS/cm
Setpoint
1000
A0009055-EN
Set value is an absolute value.
This function is primarily
suited to batch operation and
single-sided limit switches.
7.4.7 Relay configuration
For the "RELAY" function group, you require a relay card which is not available in the basic
device version.
The following relay contacts can be selected and configured as desired (max. four contacts,
depending on options installed):
Limit contactor for conductivity measured value: R2 (1)
Limit contactor for temperature: R2 (2)
PID controller: R2 (3)
Timer for cleaning function: R2 (4)
Chemoclean function: R2 (5)
USP/EP: R2 (6) and R2 (7) (for Plus Package, only conductive sensors)
Limit contactor for conductivity measured value and temperature
The transmitter has different ways of assigning a relay contact. Switch-on and switch-off
points and pickup and dropout delays can be assigned to the limit contactor. In addition,
you can configure an alarm threshold to output an error message and to start a cleaning
function in conjunction with this.
These functions can be used both for the primary value and for temperature measurement.
Please refer to the switch states in →  29 for a clear illustration of the relay contact
states.
When the measured values increase (maximum function), the relay contact is closed as
of t2 after the switch-on point (t1) has been exceeded and the pickup delay has elapsed
(t2-t1).
The alarm contact switches if the alarm threshold (t3) is reached and the alarm delay
(t4-t3) has also elapsed (errors E067 to E070).
When the measured values decrease, the alarm contact is reset when the value falls
below the alarm threshold (t5) again, as is the relay contact (t7) after the dropout delay
(t7-t6).
If the pickup and dropout delays are set to 0 s, the switch-on and switch-off points are
also switch points of the contacts.
The same settings can also be made for a minimum function in the same way as for the
maximum function.

Table of Contents

Related product manuals