Commissioning Liquisys M CLM223/253
80 Endress+Hauser
Coding Field Range of
adjustment
(factory
settings in
bold)
Display Info
C134 Calculated
installation factor
is displayed
1
0.10 to 5.00
C134
InstFact
1
A0009158-EN
The installation factor
indicates the dependency
between the measured
value and the distance
between the sensor and
pipe wall. See the Technical
Information of the sensor
used for more information.
C135 Calibration status is
displayed
o.k.
E xxx
C135
Status
o.k.
A0009159-EN
If the calibration status is
not o.k., a reason for the
error is provided on the
second line of the display.
C136 Store calibration
result?
Yes
No
New
C136
Store
yes
A0009160-EN
If C135 = E xxx, then only
No or New.
If New, return to C.
If Yes/No, return to
"Measurement".