Liquisys M CLM223/253 Commissioning
Endress+Hauser 73
Coding Field Range of
adjustment
(factory
settings in
bold)
Display Info
K7 Enter concentration
value for K6
0.00 %
0.00 to 99.99 %
K7
%
concentr.
0.00
A0009120-EN
Engineering unit as
selected in A2.
Format as selected in A3.
K8 Enter temperature value
for K6
0.0 °C
-35.0 to 250.0
°C
K8
C
°
Temp.val.
0.0
A0009121-EN
K9 Message as to whether
table status is OK
Yes
No
K9
Status ok
yes
A0009122-EN
Read only.
If "No", then set correct
values in the table (all
settings made up until
now remain intact) or go
back to the measuring
mode (then the table is
not valid).
7.4.10 Service
Coding Field Range of
adjustment
(factory settings in
bold)
Display Info
S
SERVICE
function group
S
SERVICE
A0008408-EN
Service function settings.
S1 Select language ENG = English
GER = German
FRA = French
ITA = Italian
NL = Dutch
ESP = Spanish
S1
Language
ENG
A0008409-EN
The option selected only
applies for the fault-
signaling contact, not for the
error current.
S2 Configure a hold S+C = hold during
configuration and
calibration
Cal = hold during
calibration
Setup = hold during
configuration
None = no hold
S2
Auto HOLD
S+C
A0008413-EN
S = setup
C = calibration
S3 Manual hold Off
On
S3
Man. HOLD
off
A0008414-EN
The setting is retained even
in the event of a power
failure.