Commissioning Liquisys M CLM223/253
72 Endress+Hauser
As the temperature is already processed in the concentration tables, the settings for
temperature compensation in the Setup 2 menu (fields B2 and B3) are ineffectual for
concentration measurement.
Functions marked in italics are not supported by the basic device version.
Coding Field Range of
adjustment
(factory
settings in
bold)
Display Info
K
CONCENTRATION
function group
K
CONCENTRA
A0009113-EN
Four different
concentration fields can be
entered in this function
group.
K1 Select the active
concentration curve on
which the calculation of
the display value is
based
1
1 to 4
K1
act.curve
1
A0009114-EN
The curves are
independent of one
another. Four different
curves can be defined.
K2 Select the curve to be
edited
1
1 to 4
K2
editCurve
1
A0009115-EN
If a curve is being edited,
another curve should be
used to calculate the
current display values.
Example: If curve 2 is
being edited, either curve
1, curve 3 or curve 4
should be active (see K1).
K3 Select table option Read
Edit
K3
Table
read
A0009116-EN
This option is valid for all
concentration curves.
K4 Enter number of
reference triplets
1
1 to 10
K4
No. Elem.
1
A0009117-EN
Each triplet consists of
three numeric values.
K5 Select triplet 1
1 to number of
triplets in K4
Assign
K5
Sel.Elem.
1
A0009118-EN
Any triplet can be edited.
"Assign" takes the user to
K9
K6 Enter uncompensated
conductivity value
0.0 mS/cm
0.0 to 9999
mS/cm
K6
mS/cm
conduct.
0.0
A0009119-EN
The system automatically
runs through the K5 to K6
function chain as often as
corresponds to the value
indicated in K4. The
display then goes to K9.