A0009147-EN
If the calibration status is
not o.k., a reason for the
error is provided on the
second line of the display.
C114 Store calibration
result?
Yes
No
New
A0009143-EN
The sensor should be
immersed in such a way
that there is sufficient
distance to the vessel wall
(the installation factor has
no influence if a > 15 mm).
Immerse the sensor in the calibration solution.
The following section describes calibration
with the temperature-compensated
conductivity value of the reference solution.
If calibration is to be performed with the
uncompensated conductivity value, you
must set the temperature coefficient α to
zero.
A0025598
C121 Enter calibration
temperature (MTC)
25 °C
-35.0 to 250.0 °C
Only available if B1 = fixed.
C122 Enter the α value
of the calibration
solution
2.10 %/K
0.00 to 20.00
%/K
A0009150-EN
The value is given in the
Technical Information for
all E+H calibration
solutions. You can also use
the printed-on table to
calculate the value.
Set α to 0 for calibration
with uncompensated values.
C123 Enter the correct
conductivity value
of the calibration
solution
Current
measured value
0.0 μS/cm to
9999 mS/cm