Commissioning (installation and configuration) Micropilot FMR62
16 Endress+Hauser
4.6 Parameters and default settings for SIL mode
The following parameters affect the safety function. However, they may be freely
configured in accordance with the application. In increased safety mode, it is necessary to
confirm the configured values during the remainder of the commissioning process.
Confirmation is not required in expert mode. It is recommended to note down the
configured values!
Setup
Empty calibration
Full calibration
Tank type
Only for liquids
Bin type
Only for bulk solids
Max. filling speed solid
Only for bulk solids
Max. draining speed solid
Only for bulk solids
Medium group
Advanced setup → Level → Tank/silo height
The following parameters affect the safety function and are not freely configurable in
increased safety mode. Instead, they are automatically set by the device to the safety-
oriented values mentioned at the start of SIL/WHG confirmation. If these parameters are
to be set to other values, expert mode must be selected.
Setup
Advanced setup → Level → Level correction
Default value: 0
Advanced setup → Linearization → Linearization type
Default value: None
Advanced setup → Current output 1 → Damping output
Default value: 0
Expert
Sensor → Level → Distance offset
Default value: 0
Sensor → Level → L max. drain speed
Default value: 0
Sensor → Level → L max. fill speed
Default value: 0
Sensor → Level → Level limit mode
Default value: Off
Sensor → Level → Output mode
Default value: Level linearized
Sensor → Safety settings → Jump delay echo lost
Default value: Off
Sensor → Safety settings → Delay time echo lost
Default value: 3 s
Sensor → Safety settings → Output echo lost
Default value: Alarm
Sensor → Safety settings → In safety distance
Default value: Alarm
Sensor → Echo tracking → Evaluation mode
Default value: No history
Output → Current output 1 → Turn down
Default value: Off
Output → Current output 1 → Measuring mode
Default value: Standard
Communication → Configuration → HART address
Default value: 0