Electrical connection Flowmeter Proline 200 PROFINET with Ethernet-APL
26 Endress+Hauser
The following sequence of steps is recommended :
1. Mount the sensor and transmitter.
2. Connect the .
3. Connect the transmitter.
How the connecting cable is connected in the transmitter housing depends on the
measuring device approval and the version of the connecting cable used.
In the following versions, only terminals can be used for connection in the transmitter
housing:
• Order code for "Electrical connection", option B, C, D
• Certain approvals: Ex nA, Ex ec, Ex tb and Division 1
• Use of reinforced connecting cable
In the following versions, an M12 device connector is used for connection in the
transmitter housing:
• All other approvals
• Use of connecting cable (standard)
Terminals are always used to connect the connecting cable in the sensor connection
housing (tightening torques for screws for cable strain relief: 1.2 to 1.7 Nm).
Connecting the sensor connection housing
A0034167
1. Loosen the securing clamp.
2. Unscrew the housing cover.